Eletrical Safety Analyzer (ESA) Safe Test 50 - Rigel
SPECIFICATIONS:
Earth Continuity | |
---|---|
Method | 2 wire ZAP technique. |
Test Current | >±200mADC into 2Ω |
Max Test Voltage | 4-24Vrms o/c |
Measuring Range (low range) | 0.001 – 0.999Ω / 0.001Ω resolution |
Measuring Range (mid range) | 1.00 – 9.99Ω / 0.01Ω resolution |
Measuring Range (high range) | 10.0 – 19.9Ω / 0.1Ω resolution |
Accuracy | ± 1% of value, ± 5mΩ |
Circuit Protection | Test inhibited if ≥30VAC or DC at 4mm inputs |